Motion Accuracy of Pneumatic Stepper Motor-Driven Robotic System Developed for MRI-Guided High-Intensity Focused Ultrasound Treatment of Prostate Disease

Table 1

SNR computed from the MRI images obtained under the three conditions.

ConditionMean intensitySDSNRNormalized SNR (%)

Phantom only2,936.824.3121.0100.0
Robot connected2,935.224.9118.097.5
Robot in operation2,931.424.9117.797.2